Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
116
SUEZ © 2022
The Sample Inlet Block is designed for Online sampling only. Sampling from a vial is not
possible with this configuration
.
NOTE:
The Sample Inlet Block is designed for Online sampling
only. Sampling from a vial is not possible with this configuration.
S
TEAM
E
XPOSURE
The iOS is rated at IP 55 and can withstand short-term steam exposure of up to 40 psi
steam for 30 minutes. When exposed to steam, the Analyzer should be powered off or
in standby mode, and not in analysis mode. After exposure to steam, allow the
Analyzer to cool before returning to Analysis mode. Observe the following cool-down
intervals before starting analysis after steam exposure:
Warning
The Sample Inlet Block may be hot. If the sample water is hot, be careful
to avoiding touching the Sample Inlet Block.
Ambient Temperature
Time Interval
Below 20 °C
15 minutes
20 - 30 °C
30 minutes
30 - 40 °C
90 minutes