L
OW
L
EVEL
A
CCURACY
& P
RECISION
V
ERIFICATION
P
ROTOCOL
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
228
SUEZ © 2022
5.3.6 When analysis of the vial is complete, remove it from the iOS and click
Vial removed -> Continue
.
5.4 The
Low Level Accuracy & Precision Verification: Protocol Complete
screen
appears with the
Summary
tab active:
•
Vial 1–2 — Select a tab to view analysis details for each
standard.
•
Summary — Select this tab to view the summary of analysis, including a
passed
or
failed
result. It also contains
,
Export
, and
Close
buttons.
5.5 Referencing the
Summary
tab, record the results data on the
Accuracy &
Precision Verification Worksheet
.
5.6 If the verification failed, leave the screen as is and contact Technical Support
or your local service provider. Before calling, e-mail the backup and system
settings files (created in advance of running calibration protocol). After
analysis of the data, you may be advised to pursue troubleshooting steps.
5.7 To print or export the results, click the corresponding button.