![Suez Sievers M500e Operation And Maintenance Manual Download Page 303](http://html1.mh-extra.com/html/suez/sievers-m500e/sievers-m500e_operation-and-maintenance-manual_1389721303.webp)
W
ARNING
AND
E
RROR
D
ESCRIPTIONS
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
303
SUEZ © 2022
10013
Warning
Autozero
Scheduled TOC autozero offset exceeded limit.
Occurs when scheduled autozero fails
11000
Error
Communication
Communication error, could not open connection.
Occurs when establishing connection between the GUI and the
Instrument Controller, if the serial port cannot be opened for
communication
11001
Error
Communication
Communication error, read timeout.
Occurs when the GUI cannot read from the serial port because the
Instrument Controller closed the port.
11002
Error
Communication
Communication error, write timeout.
Occurs when the GUI cannot write to the serial port because the
Instrument Controller closed the port.
11003
Error
Communication
Communication error, ill formed command.
Occurs during Modbus communication, if invalid response is
encountered.
11400
Error
Datastore
Datastore error, invalid file path.
Occurs when records cannot be saved or read because data
directory does not exist or the datastore file path is invalid. Also if
during data export, the export file path is invalid (directory does not
exist or does not allow writing).
11401
Error
Datastore
Datastore error, could not open file for read/write.
Occurs when records cannot be saved because a datastore file
cannot be created or opened for read/write.
11402
Error
Datastore
Datastore error, could not open file for writing.
Occurs when records cannot be saved because a datastore file
cannot be created or opened for writing.
11403
Error
Datastore
Datastore error, could not create file.
When records cannot be saved because a datastore file cannot be
created.
11404
Error
Datastore
Datastore error, record encryption failed.
Occurs when a datastore record cannot be saved because
encryption failed.
11405
Error
Datastore
Datastore error, could not add new record.
Occurs when a new datastore record cannot be saved because the
maximum count for the period was reached or the timestamp is
invalid.
11406
Error
Datastore
Datastore error, could not add new record.
Occurs when datastore records cannot be read (e.g., when
displaying, exporting, printing).
#
Warning/
Error
Name
Description