3 Design and functional description
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
73
3.2.8
Status, alarm and info display
Fig. 30: Status, alarm, and info display with active alarm
1 Alarm display
4 Trigger display
2 Alarm suppression time display
5 Nebulization time display
3 Pre-oxygenation time display
6 Alarm history icon
Fig. 31: Status, alarm, and info display, alarm history display
The alarm display always shows the active alarm with the highest priority
(notification = blue, MP alarm = yellow, HP alarm = red). Pressing the
alarm display opens the alarm history (see chapter 3.2.3.3). If there are
no active alarms, but a non-acknowledged alarm history exists, the
corresponding icon appears on the alarm display. The icon disappears
after deleting the alarm history.
Pressing the »Alarm suppression« button (see chapter 3.1.2) mutes all
alarms for a duration of two minutes. The corresponding icon appears on
the display, and a countdown timer indicates the remaining alarm
suppression time.
Pressing the »Preoxy« button (see chapter 3.1.3) displays the
corresponding icon, and the preset preoxygenation time starts to count
down.
Pressing the »Aerosol« button (see chapter 3.1.3) displays the
corresponding icon, and the preset nebulization time starts to count down.
Alarm display and
alarm history
Alarm suppression
display
Preoxygenation display
Nebulization display
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...