3 Design and functional description
44
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
NOTE
For safety reasons, the volume-controlled ventilation modes cannot be selected in the
mode for preterm infants and newborns.
Pressing this key will start ventilation using the standard parameters for
children (see chapter 6.6).
Pressing this key will start ventilation using the standard parameters for
adults (see chapter 6.6).
Pressing this key will suppress acoustic alarms for 2 min. If the key is
pressed again within 5 s, the alarm suppression is canceled again.
Pressing the key again after 5 s has passed will start a new 2 minute
countdown.
NOTE
During therapeutic procedures, the acoustic alarm can be completely suppressed for
2 min by pressing the alarm suppression button before the occurrence of the first
alarm.
Child fast tracking key
Adult fast tracking key
Acoustic alarm
suppression