3 Design and functional description
60
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
NOTE
Switching off the device completely will automatically delete all entries from the
alarm history.
NOTE
The alarm history is retained in the event of a power outage of less than 30 s. In this
case, energy is supplied by the internal battery.
NOTE
In case of a total power failure, all entries are deleted from the alarm history.
Pressing the »Limits 1-3« fields lets you view all the alarm limits and
adjust them to the patient's needs. If limits are breached during the
current ventilation, the corresponding parameter field turns yellow or red,
depending on the alarm priority.
Fig. 19: Alarms sub-menu
1 Name and unit
3 Current measured value
2 Upper limit
4 Lower limit
Limits