8 CO2 measurement (optional)
178
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
WARNING
When connecting the IRMA™ probe to the system of an infant, direct contact of the
probe with the infant's body must ALWAYS be prevented. If the probe comes in
contact with the infant's body for whatever reason, insulating material must be used.
CAUTION
Do not use the IRMA™ probe outside the specified ambient temperature.
CAUTION
Do not expose the cable of the IRMA™ probe to tensile forces.
8.1.4
Installing the CO
2
measuring probe
To install the CO
2
measuring probe:
Connect the IRMA™ connecting cable to the CO
2
sensor input on
the right side of the
EVE
TR
(see Fig. 39). Switch on the
EVE
TR
.
Before connecting the IRMA™ airway adapter to the patient system,
the gas values and curves on the
EVE
TR
monitor must be checked.
Fig. 99: Airway adapter for adults/children and infants
Secure the IRMA™ probe to the IRMA™ airway adapter. The probe
audibly locks into place when secured properly.