11 Care and maintenance
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
227
System components Recommended
treatment intervals
Automated
thermal
treatment
Manual
treatment
Sterilization
Pressure measuring
adapter NCPAP
After every patient/
weekly
Yes
Yes
Yes
SpO
2
sensor
After every patient/
weekly
No
Surfaces
No
Observe the manufacturer’s operating
manual!
CO
2
sensor
After every patient/
weekly
No
Surfaces
No
Observe the manufacturer’s operating
manual!
Mobile stand,
brackets
After every patient/
weekly
No
Surfaces
No
Tab. 48: Treatment table
11.9
Safety checks
Safety checks must be carried out annually by the manufacturer or the
authorized F
RITZ
S
TEPHAN
G
MB
H
customer service team.
11.10
Maintenance
For reasons of device safety, we recommend performing basic
maintenance on the
EVE
TR
annually in connection with the safety
checks. All maintenance
Basic maintenance (annually)
Major maintenance (every five years)
must be performed according to maintenance instructions by the
authorized F
RITZ
S
TEPHAN
G
MB
H
customer service team.