3 Design and functional description
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
61
WARNING
The alarm limits must be verified by medical personnel and possibly adapted to the
current patient situation. The alarm limits must always be based on the patient's needs.
Using extreme settings that are not medically indicated may render the alarm system
useless and put the patient at risk.
The
EVE
TR
always starts with the following preset alarm limits:
Parameter
Unit
Lower limit
Upper limit
PAW
mbar
-
Pinsp+10
(maximum 60 mbar)
PEEP
mbar
Set value -2-
Set value + 2
VTe
(VC mode)
ml
Set value
VT −30%
Set value
VT +30%
VTe
(PC mode)
ml
Adult:
350
Adult:
650
Child:
140
Child:
260
Preterm infants
and newborns:
14
Preterm infants and
newborns:
26
MVe
(VC mode)
ml
(Set values VT·f)
−30%
(Set values VT·f)
+30%
MVe
(PC mode)
ml
Adult:
4.2
Adult:
7.8
Child:
2.8
Child:
5.2
Preterm infants
and newborns:
0.42
Preterm infants and
newborns:
0.78
fspon
1/min
-
50
Apnea
s
-
15
EtCO
2
mmHg
Vol.%
kPa
30
4
4
45
6
6
O
2
%
Set value −10%
minimum
21%
Set value +10%
maximum 100%
Tab. 6: Preset alarm limits
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...