CAUTION
The measurement of the minute volume (MV) may be affected by the medication
nebulization.
CAUTION
Remove the medication nebulizer after use. Unintended nebulization may put the
patient at risk and impair breathing.
CAUTION
The
EVE
TR
features built-in automatic compensation. As a result, the Aerosol button
(see chapter 3.1.3) should be pressed only when a medication nebulizer is connected.
Otherwise, the resulting breathing volume is too low.
NOTE
Aerosol nebulization is only possible if O
2
is connected to the ventilator.
NOTE
Aerosol nebulization is automatically synchronized with inspiration.
NOTE
In volume-controlled mode, the volume is automatically compensated for during
nebulization.
For medication nebulization, install the nebulizer as shown in Fig. 71 and
fill according to the operating manual provided with the nebulizer.