10 Troubleshooting
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
203
Leakage
(Alarm delay: 3 s)
HP
PA
If the inspiratory
tidal volume is 80%
more than the
expiratory tidal
volume after three
breaths
Check the patient's
condition
Check the tube system
Check the ETT
Fan failure
MP
TA
Fan malfunction
Check the fan, replace if
necessary
Low O
2
pressure
MP
TA
Oxygen delivery
pressure from the
gas cylinder or
central gas supply
too low (< 2.7 bar)
Check the oxygen
delivery pressure and
replace the O
2
cylinder if
necessary
Check the oxygen
pressure in CGS
O
2
high
(Alarm delay: 30s)
MP
PA
Calculated O
2
value
is above the
automatic alarm
limit value
Compare the set O
2
value
with the calculated value
In case of larger
deviations, contact F
RITZ
S
TEPHAN
G
MB
H
customer service
O
2
low
(Alarm delay: 30s)
MP
PA
Calculated O
2
value
is below the
automatic alarm
limit value
Compare the set O
2
value
with the calculated value
In case of larger
deviations, contact F
RITZ
S
TEPHAN
G
MB
H
customer service
Parameter error
MP
TA
Internal
communication
error
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...