12 Electromagnetic emissions and immunity
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
235
12
Electromagnetic emissions and
immunity
NOTE
Electrical medical devices are subject to special precautions regarding
electromagnetic compatibility (EMC) and must be installed and put into service
according to the EMC information provided in the accompanying documentation.
NOTE
Portable and mobile RF communications equipment can affect electrical medical
devices.
12.1
Electromagnetic emissions
The
EVE
ventilator is intended for use in the electromagnetic
environment specified below. The customer or user of the
EVE
ventilator
should ensure that it is used in such an environment.
Emissions
Compliance
Electromagnetic environment guideline
RF emissions as defined in
CISPR 11
Group 1
The
EVE
ventilator uses RF energy for its internal
function only. Therefore, its RF emissions are very
low and are not likely to cause any interference with
nearby electronic equipment.
RF emissions as defined in
CISPR 11
Class A
The
EVE
is suitable for use in establishments other
than homes and those directly connected to the public
low-voltage power supply network that also supplies
buildings used for domestic purposes.
Harmonics as defined in
IEC 61000-3-2
Class B
Voltage fluctuations/flicker
as defined in IEC 61000-3-3
Met
Tab. 50: Electromagnetic emissions (IEC 60601-1-2)
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...