3 Design and functional description
62
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
To change the alarm limits, select the corresponding upper or lower limit
on the touchscreen. The field opens and is illuminated yellow. Use the
control knob to adjust the value within the preset limits.
Parameter
Unit
Lower limit Upper limit Resolution
PAW
mbar
-
11 – 60
1
PEEP
mbar
0 – 29
1 – 30
1
VTe
ml
0 – 100
1 – 100
1
100 – 490
100 – 500
5
490 – 990
500 – 2000
10
EtCO
2
Vol.%
0.0 – 11.9
0.1 – 12.0
0.1
mmHg
0 – 89
1 – 90
1
kPa
0 – 11.9
0.1 – 12.0
0.1
PI
%
Off,
0.03 – 0.1
0.04 – 0.1
0.01
0.1 – 1
0.1 – 1
0.1
1 – 18
1 – 19, Off
1
PVI
%
1 – 98/Off
2 – 99/Off
1
SpO
2
%
88 – 98
91 – 99/Off
1
SpCO
%
1 – 97
2 – 98/Off
1
fspon
1/min
5 – 120/Off
1
Apnea
s
4 – 60
1
Pulse
bpm
30 – 230
35 – 235/Off 1
Tab. 7: Adjustable alarm limits
Pressing the »Auto« field sets the PAW, MVe, VTe, EtCO
2
and PEEP
limits to the system's automatic alarm limit presets.
Changing alarm limits
Auto