1 General information
26
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Parameter
Meaning
Setting range
Resolution
I:E
Inspiration/
expiration ratio
1:150, 150:1
1
f
Frequency
1 – 150 1/min
1
Apnea
Apnea duration
1 – 60 s
1
O
2
O
2
ratio of breathing air 21 – 100%
1
Trigger
Flow trigger
0.2 – 15 l/min
0.1
Ramp time
Ramp time
0.06 – 30 s
0.01
Flow
Flow O
2
therapy
2 – 20 l/min
1
ETS
Expiration trigger
sensitivity
5 – 70%
1
Tube
compensation
Intensity of tubus
compensation
0 – 100%
1
Tube Ø
Tube diameter
2 – 12 mm
0.5
*
Ensured by using a redundant pressure measurement and limitation
Measured value Unit
Resolution
Display range
Pressure
(accuracy
1
:
P =
2 mbar + 4% of displayed value)
Ppeak
mbar
1
−20 – 99
Pplat
mbar
1
−20 – 99
Pmean
mbar
1
−20 – 99
PEEP
mbar
1
−20 – 99
Flow
(STPD, accuracy
V` =
15%)
V´min
l/min
0.1
−200 – 200
V´max
l/min
0.1
−200 – 200
Volume
𝑉 = ∫ 𝑉´(𝑡)
(STPD, accuracy
V =
4.0 ml +15%)
VTe
ml
1
0 – 3000
VTspon
ml
1
0 – 3000
VTleak
ml
1
0 – 100
MVe
l/min
0.01
0 – 999
MVspon
l/min
0.01
0 – 999
1
Accuracy assumes an ambient temperature of 0 – 40° C.
Monitoring
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...