7 Ventilation modes
154
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Ventilation mode Adjustable parameters
PC-SIMV
with PRVC
O
2
Inspiration time (Tinsp) and breathing frequency (f)
(the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Positive end expiratory pressure (PEEP)
Ramp time
Target volume (VT)
Tab. 24: PC-SIMV with PRVC
Ventilation mode Adjustable parameters
PC-SIMV
with PSV
O
2
Inspiration time (Tinsp) and breathing frequency (f)
(the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Inspiration pressure (Pinsp)
Positive end expiratory pressure (PEEP)
Ramp time
Pressure support (
Psupp)
(relative set value with reference to PEEP)
Trigger
Expiratory trigger sensitivity (ETS)
Tab. 25: PC-SIMV with PSV