10 Troubleshooting
208
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
10.2
Selftest error
Error during
power supply test
HW
Battery malfunction
Battery empty
Check battery level, charge
battery if necessary
If the error cannot be
corrected, contact
F
RITZ
S
TEPHAN
G
MB
H
customer service, providing
the error code displayed
Error during valve
test
HW
Valve malfunction
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service, providing
the error code displayed
Error during sensor
test
HW
Sensor system malfunction
Check secure connection of
the patient tube system
If the error cannot be
corrected, contact
F
RITZ
S
TEPHAN
G
MB
H
customer service, providing
the error code displayed
Error during
turbine test
HW
Turbine malfunction
Check secure connection of
the patient tube system
Check system for leakage
Check sealing plug
on Y piece
If the error cannot be
corrected, contact
F
RITZ
S
TEPHAN
G
MB
H
customer service, providing
the error code displayed
Tab. 45: Selftest error