7 Ventilation modes
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
139
7
Ventilation modes
7.1
Invasive and non-invasive ventilation modes
The
EVE
TR
provides both invasive and non-invasive (NIV) ventilation
modes. Whereas all volume-controlled ventilation modes
(VC-CMV, VC-SIMV, see chapter 7.3.1) are only suitable for invasive
ventilation, the pressure-controlled ventilation modes (PC-SIMV,
PC-CMV, PC-ACV, PC-ACV+ (optional), CPAP, and DUOPAP, see
chapter 7.3.2) can be used for non-invasive ventilation.
The non-invasive ventilation modes nCPAP, nPC-SIMV, nPC-ACV,
nPC-ACV+, nDUOPAP, and nPC-CMV are used for the ventilation of
preterm infants and newborns. In this case, ventilation is carried out
without a flow sensor because pressure measurement is performed
proximally via an adapter.
NOTE
The ventilation modes for preterm infants and newborns, as well as for PC-ACV+ and
nPC-ACV+, are not included in the scope of supply and available only with a
corresponding license (see chapter 1.2.1).
A flow limit is available in NEO mode for non-invasive ventilation
modes. The preset flow limit is 9 l/min. It can be set within a range of
5-30 l/min in the corresponding field or switched off completely.
Fig. 82:Flow limit