11 Care and maintenance
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
217
CAUTION
Do not use abrasive cleaners or brushes on the touchscreen. They can scratch the
display.
Prior to cleaning the touchscreen during ongoing ventilation, ALWAYS press the
Lock Touchscreen button (see chapter 3.1.3).
All disinfectant residue must be completely removed afterwards.
Suitable cleaning agents with regard to material compatibility are
aldehyde-free quick disinfecting wipes, such as Bacillol
®
wipes or tissues
made by Paul Hartmann AG, Heidenheim or Mikrozid
®
PAA wipes
made by
Schülke, Norderstedt.
11.6
Sterilization
Components labeled in this operating manual as suitable for sterilization
can be sterilized with hot steam at temperatures of up to 134° C.
NOTE
Only use vacuum steam sterilizers!
Sterilizers with a fractionated vacuum are preferable.
The hot steam sterilizer must be loaded according to the manufacturer’s
instructions for use and the corresponding program started.
Temperature
Hold/dwell time
134° C
3 – 18 min.
Tab. 47: Standard procedure for steam sterilization*
*All standard steam sterilization procedures can be used.
Disinfectants