6 Operation
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
131
6.7
Ventilation tube system
Standard tube systems are preconfigured for the various patient groups. If
other tube systems are used, make the appropriate settings in this menu or
in the System Settings/Sensors/Tube system menu (see chapter 4.2.4).
The corresponding tube system must be selected for this based on the
type designation or a pictogram.
If no new patient is displayed, the presetting for the tube system of the
last ventilation is adopted.
Fig. 76: Presetting for the patient tube system
ATTENTION
F
RITZ
S
TEPHAN
G
MB
H explicitly recommends that only patient tube systems included
in the list of accessories are used. Otherwise, ventilator function may be impaired!
If it is necessary to use a non-standardized tube system that has not been tested by
F
RITZ
S
TEPHAN
G
MB
H, select the "Other Tube System" option. Then enter the tube
system's parameters manually.
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...