10 Troubleshooting
206
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Blower
failure 3
HP
TA
Internal pressure
sensor malfunction
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Blower
failure 5
HP
TA
CAN bus
communication fault
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Blower
failure 6
HP
TA
Malfunction of the
internal voltage
supply
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Disconnection
HP
PA
If the "PAW low"
limit is not breached
during the entire
breathing cycle, an
alarm is issued after
a 20 s delay
Tube system leakage
or disconnection
Check the pressure alarm
limits
Check the patient tube
system for disconnected
parts
Check that the manual
settings are correct and
plausible, adjust if
necessary
EEPROM failure
HP
TA
EEPROM
malfunction
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Measurement system
error
HP
TA
Internal measuring
error
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
MVe high
HP
PA
Expiratory breathing
volume above the
upper alarm limit
setting
Flow sensor
malfunction
Check the patient's
condition
Check minute volume
alarm limits
Check that the manual
settings are correct,
adjust if necessary
Check the flow sensor
MVe low
HP
PA
Expiratory breathing
volume below the
lower alarm limit
setting
Leak
Flow sensor
malfunction
Check the patient's
condition
Check minute volume
alarm limits
Check patient tube
system and connections
for leakage
Check that the manual
settings are correct,
adjust if necessary
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...