10 Troubleshooting
202
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
No CO
2
sampling line
MP
TA
Sampling line not
connected
Connect the sampling
line
CO
2
sensor zeroing
required
MP
TA
CO
2
calibration
failed
Repeat CO
2
calibration
Check the CO
2
adapter
and clean if necessary
Replace the CO
2
adapter,
if necessary
Replace the CO
2
sensor,
if necessary
CO
2
sensor reinsert
MP
TA
CO
2
measurement
faulty
Check the CO
2
sensor
Reinsert the CO
2
sensor
EtCO
2
high
MP
PA
Measured EtCO
2
above the upper
limit setting
Check the patient's
condition
Check the EtCO
2
upper
limit and adjust if
necessary
Check the ventilation
settings
Calibrate the CO
2
sensor,
if necessary
Clean the CO
2
sensor/
cuvette, if necessary
EtCO
2
low
MP
PA
Measured EtCO
2
below the lower
limit setting
Check the patient's
condition
Check the EtCO
2
lower
limit and adjust if
necessary
Check the ventilation
settings
Calibrate the CO
2
sensor,
if necessary
Clean the CO
2
sensor/
cuvette, if necessary
CO
2
sensor error
MP
TA
CO
2
sensor is not
detected
CO
2
sensor or
adapter
contaminated.
CO
2
sensor possibly
incorrectly attached
to the cuvette
Check the CO
2
sensor
Replace the CO
2
sensor
if necessary
Use external CO
2
monitoring if necessary
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...