11 Care and maintenance
212
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
CAUTION
Do not immerse the device in liquids! Use cleaning solutions sparingly. Excess liquid
may enter the housing and damage the device.
11.2
Information about cleaning agents and
disinfectants
CAUTION
When using cleaning agents and disinfectants, pay attention to the correct
concentration and dwell time to rule out any damage to the materials.
Always closely follow the instructions of the cleaning agent manufacturer for using
the cleaning agent!
CAUTION
When using agents other than those specified, please contact the manufacturer of the
disinfectant to confirm its compatibility with ventilation and inhalation systems
(safety data sheet, toxicity).
CAUTION
When using cleaning agents and disinfectants, observe the rules established by the
professional associations on the use of cleaning agents and disinfectants.
CAUTION
Do not use petroleum-based cleaning agents, acetone, or other harsh cleaning agents!
These agents can damage the device material and cause malfunctions.
CAUTION
Disinfectants based on amines and their derivatives can damage silicone parts
(e.g. patient tubes) and are therefore not suitable for use on the device.