7 Ventilation modes
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
155
Non-invasive pressure-controlled synchronized intermittent
7.3.2.4
mandatory ventilation (nPC-SIMV)
The nPC-SIMV corresponds to the behavior of the PC-SIMV. It is only
available in non-invasive Neo mode. The device is operated without a
flow sensor (PNT) in this mode. The pressure is measured proximally via
the pressure measurement adapter set.
Ventilation mode Adjustable parameters
nPC-SIMV
O
2
Inspiration time (Tinsp) and breathing frequency (f)
(the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Positive end expiratory pressure (PEEP)
Ramp time
Inspiration pressure (Pinsp)
Trigger
Flow limitation (in NEO mode only)
Tab. 26: nPC-SIMV
Ventilation mode Adjustable parameters
nPC-SIMV
with PSV
O
2
Inspiration time (Tinsp) and breathing frequency (f)
(the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Inspiration pressure (Pinsp)
Positive end expiratory pressure (PEEP)
Ramp time
Pressure support (
Psupp)
(relative set value with reference to PEEP)
Trigger
Expiratory trigger sensitivity (ETS)
Tab. 27: nPC-SIMV with PSV