8 CO2 measurement (optional)
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
191
8.2.7
Cleaning the CO
2
analyzer
The ISA™ CO
2
analyzer must be cleaned regularly. Use a cloth
moistened with alcohol or isopropyl alcohol (< 70%) for cleaning.
While cleaning the analyzer, the Nomoline sample collection tube must
remain connected to prevent cleaning liquid or dust from entering the
analyzer via the CO
2
connection.
CAUTION
Do not sterilize the ISA™ CO
2
analyzer or submerge it in liquids.
CAUTION
The Nomoline sample collection tubes are not sterile. To prevent damage, do not
autoclave any part of the sample collection tube.