
®
7 Ventilation modes
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
157
NOTE
The PC-ACV ventilation mode can be used with volume guarantee (PRVC) (see
chapter 7.4.1). It also includes an option for tube compensation (see chapter 7.4.3).
Ventilation mode Adjustable parameters
PC-ACV
with PRVC
O
2
Inspiration time (Tinsp) and breathing frequency (f)
(the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Positive end expiratory pressure (PEEP)
Ramp time
Target volume (VT)
Trigger
Tab. 29: PC-ACV with PRVC
Non-invasive pressure-controlled assist/control ventilation (nPC-
7.3.2.6
ACV)
The nPC-ACV corresponds to the behavior of the PC-ACV. It is only
available in non-invasive Neo mode. The device is operated without a
flow sensor (PNT) in this mode. The pressure is measured proximally via
the pressure measurement adapter set.
Ventilation mode Adjustable parameters
nPC-ACV
O
2
Inspiration time (Tinsp) and breathing frequency
(f)
(the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Positive end expiratory pressure (PEEP)
Ramp time
Trigger
Flow
limitation
(in NEO mode only)
Tab. 30: nPC-ACV
Summary of Contents for EVETR
Page 1: ...EVETR Transport Ventilator Instructions for use...
Page 32: ......
Page 40: ......
Page 86: ......
Page 106: ......
Page 124: ......
Page 192: ......
Page 210: ......
Page 234: ......
Page 240: ......
Page 246: ......
Page 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...