7 Ventilation modes
156
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Pressure-controlled assist/control ventilation (PC-ACV)
7.3.2.5
Like PC-SIMV, the assist/control mode is a synchronized ventilation
mode. In PC-ACV mode, however, the ventilator supports all of the
patient’s own breathing efforts that exceed the preset trigger threshold
with a mechanical breath. This triggers the patient's mandatory
inspiration, but its characteristic is controlled by the device. The patient
cannot trigger a new mandatory breath until this time has elapsed.
Fig. 87: PC-ACV
Ventilation mode Adjustable parameters
PC-ACV
O
2
Inspiration time (Tinsp) and breathing frequency
(f) (the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Inspiration pressure (Pinsp)
Positive end expiratory pressure (PEEP)
Ramp time
Trigger
Tab. 28: PC-ACV
NOTE