8 CO2 measurement (optional)
176
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
WARNING
The IRMA™ airway adapter is a disposable product and not designed to be reused.
Reuse of the disposable adapter may cause cross-infection.
WARNING
Do not use the IRMA™ airway adapter for adults and children with infants because
the adapter increases the dead space in the patient's breathing system by 6 ml.
WARNING
Do not use the IRMA™ airway adapter for infants with adults because this may cause
excessive flow resistance.
WARNING
Mobile and RF communication devices may impair the measurements. It must be
ensured that the probe is used in an electromagnetic environment as specified in the
operating instructions.
WARNING
The use of RF electrosurgical devices in the vicinity of the IRMA™ CO
2
analyzer or
the
EVE
TR
may cause interference and erroneous measurements.
WARNING
The IRMA™ CO
2
analyzer is not designed for use in MRI environments.
WARNING
Do not use the IRMA™ airway adapter together with metered-dose inhalers or
medication nebulizers because this impairs the translucency of the adapter's view
ports.