MG40 Series
Area of use
Th
is function is used to set and acquire the area where the MG40 series is to be used.
If it is not set, it will not be possible to transfer from the setup mode to the measuring mode.
When STD2 serves as the Area of use setting, the value in parentheses ( ) is used as the output resolution
setting.
Th
e scale diff ers between JPN/STD1 and STD2. When the setting has been established for an area with a
diff erent scale, the setting values related to the output resolution will be restored to the factory statuses.
Set
Setup
mode
Ethernet
CC-Link
The command sets the area where the MG40 series is to be used.
Command format
CTR=<Value>
<Value>
0 : Not set (Factory setting)
1 : JPN (Th
is setting must be used if the MG40 series
is to be used in Japan.)
2 : STD1
3 : STD2
Return format
Execution results
Target
MG41 main unit
Save setting values
Setting values saved.
Examples of use
Transmission :
CTR=1
Return
:
OK000
Compatible command
None
Acquire
Setup
mode
Ethernet
CC-Link
Measurement
mode
The command acquires the setting for the area where the MG40 series is to be used.
Command format
CTR?
Return format
CTR=<Value>
<Value>
0 : Not set
1 : JPN
2 : STD1
3 : STD2
Target
MG41 main unit
Examples of use
Transmission :
CTR?
Return
:
CTR=2
Compatible command
None
Summary of Contents for MG40 Series
Page 2: ...MG40 Series ...
Page 4: ...ii J MG40 Series ...
Page 8: ...1 4 J MG40 Series ...
Page 20: ...3 4 J MG40 Series ...
Page 68: ...5 6 J MG40 Series ...
Page 72: ...ii E MG40 Series ...
Page 76: ...1 4 E MG40 Series ...
Page 88: ...3 4 E MG40 Series ...
Page 136: ...5 6 E MG40 Series ...