13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.1 General information on unit adjustment and calibration
Installation Manual ORTHOPHOS SL
"Adjustment of Ceph secondary
diaphragm"
(Ceph - adjustment)
In most cases the ceph secondary diaphragm can be performed directly
via precision adjustment. Only in exceptional cases, e.g. if the exposed
image area is completely outside the image section in an image acquired
with the
"Fine adjustment"
setting (E), is it necessary to perform coarse
adjustment prior to precision adjustment (F).
On the exposure with coarse adjustment (F), the exposed area is still
visible in the image field. Even in this extreme case, an adjustment would
still be possible.