13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
A
Adjustment OK
● The exposed diaphragm area must lie
centered and straight in the image
section as well as inside the
superimposed auxiliary lines.
● A surrounding white border must be
present.
● The maximum density must lie in the
center of the diaphragm area.
B
Adjustment not OK
Summary of Contents for ORTHOPHOS SL 2D