64 95 142 D3632
102
D3632
.
031.01.02
.
02 03.2016
10 Installation: Ceph arm
Sirona Dental Systems GmbH
10.5 Final installation work
Installation Manual ORTHOPHOS SL
10.5
Final installation work
The following steps apply to both left-handed and right-handed arms.
1. Use the spirit level to check that the secondary diaphragm is vertical.
To ensure this, place the spirit level on the side of the secondary
diaphragm.
2. NOTICE! Do not damage the unit. Do not move the secondary
diaphragm or sensor adapter by hand or apply any force.
Remove the top cover.
3. Remove the protective canvas (A).
4. Loosen the stud screw (B) (2-3 turns).
5. Set the tilt of the ceph arm on the screw (C). The secondary
diaphragm must stand upright. To ensure this, place the spirit level on
the side of the secondary diaphragm.
6. Retighten the stud screw firmly.
7. Attach the cover with the 3 screws (D).
A
IMPORTANT
Only perform the following steps if the secondary diaphragm is not
included in the batch.
B
C
3x M 4x35
D
Summary of Contents for ORTHOPHOS SL 2D