64 95 142 D3632
202
D3632
.
031.01.02
.
02 03.2016
13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.1 General information on unit adjustment and calibration
Installation Manual ORTHOPHOS SL
Validity of existing adjustment and calibration
The elements of the structure tree used for adjusting (calibrating) the unit
are prefixed by status indicators indicating the current status of the
corresponding adjustment or calibration operation.
Adjustment
Calibration
Not checked Invalid data record or no
record present
Adjustment necessary
Checked
Valid data record,
adjustment performed and
saved (up to date)
No adjustment necessary
Green
Valid data record;
calibration is in progress.
No calibration required
Orange
Data record available,
but not yet
saved
Data record must be saved
Red
Invalid data record or no
record present
Calibration required