13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1 General information on unit adjustment and calibration
båÖäáëÜ
13.1
General information on unit adjustment and
calibration
Please adhere to the following order when adjusting and calibrating the
system:
2D adjustment (panoramic)
●
"Sensor adjustment"
●
"Diaphragm adjustment"
●
"Symmetrie adjustment"
●
"Sensor calibration (DCS)"
If a cephalometer is installed:
●
"Adjustment of Ceph primary diaphragm"
●
"Adjustment of Ceph secondary diaphragm"
●
"Adjustment of Ceph secondary diaphragm (Quickshot)"
●
"Adjustment of Ceph main beam direction"
●
"Adjustment of earplug alignment"
3D adjustment/calibration (volume)
●
"Sensor adjustment"
●
"Diaphragm adjustment"
●
"Sensor calibration"
●
"Diaphragm calibration"
●
"Geometry calibration"
●
"Dose measurement"
Tip: It may be helpful to use the coloring function on the calibration menu
to evaluate exposures.