13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.4
Pan sensor calibration (DCS)
NOTE: For sensor calibration, 7 exposures are taken (for the different kV/
mA combinations). The exposure is only displayed in the exposure
window at the end of the calibration procedure.
1. Open the calibration menu [ → 195].
2. In the structure tree under
"Pan"
, click on the
"Sensor calibration
(DCS)"
element (S010.10).
3. Establish [ → 209] receptivity.
4. Take the exposure for the first kV/mA combination (60 kV / 5 mA ;
After taking the exposure, the unit automatically re-establishes
receptivity.
5. Take the exposure for the second Kv/mA combination [ → 209].
After taking the exposure, the unit automatically re-establishes
receptivity.
IMPORTANT
For the
"Sensor calibration (DCS)"
, a total of 7 exposures must be taken
(for the applicable kV/mA combinations). After taking the first exposure,
receptivity is automatically established again by the unit. This is
repeated until the last exposure has been taken.