64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
233
Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.2 Adjustment and calibration via the calibration menu
båÖäáëÜ
13.2.1.5
Adjustment of the ceph primary diaphragm
13.2.1.5.1
Automatic adjustment: Ceph - Primary diaphragm (standard)
Phantom for Ceph exposure
✔
The needle phantom is
not
in the bite block holder on the unit.
✔
The ceph test phantom connects to the cover of the sensor
✔
The ceph sensor is plugged into the ceph slot on the unit.
✔
The ear plug holders on the ceph arm have been moved completely
apart and swung out of the beam direction (ap).
1. In the structure tree, under
"Ceph"
, click on the
"Adjustment of Ceph
primary diaphragm"
element (S010.3).
The
"Adjustment of Ceph primary diaphragm"
menu is displayed
in the action area.
2. Establish [ → 209] receptivity.
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......