64 95 142 D3632
74
D3632
.
031.01.02
.
02 03.2016
7 Installation: Remote control
Sirona Dental Systems GmbH
7.3 Installation
Installation Manual ORTHOPHOS SL
surface installation
➢ Fasten the chassis firmly to the wall with the three screws (C).
Shortening the cable
1. Shorten cable L117 to the desired length.
2. Expose the cable shield (D) (see condition on delivery). The length of
the wires should be approx. 60 mm (2 3/8").
3. Place the shortened shielding over the insulation and wrap the
shielding with 3 layers of self-adhesive copper foil.
4. Strip the wire ends to 5 mm (1/5").
5. Crimp on the end sleeves. Crimp the orange with white/blue and blue
with white/orange wires each into an end sleeve.
3x 4x30
C
10mm
3/8”
ca. 250mm
10”
60mm
2 3/8”
D
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......