13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1 General information on unit adjustment and calibration
båÖäáëÜ
13.1.2
Calibration menu
The menu guides you through the procedure to adjust and calibrate the
unit.
13.1.2.1
Opening the calibration menu
✔
IT package is installed.
✔
is installed.
1. Start SIDEXIS 4.
2. In SIDEXIS 4, call the
"ORTHOPHOS SL"
/
"Device administration"
configuration menu.
Configuration menu "ORTHOPHOS SL" /"Device administration"
3. Select the desired X-ray component from the unit list.
4. Click on the
"Device calibration"
button.
A dialog box to enter the service password opens.
5. Enter the service password (see section "Password protection") and
confirm by pressing
"OK"
.
The calibration menu opens.