64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
49
Sirona Dental Systems GmbH
5 Installation: X-ray unit
Installation Manual ORTHOPHOS SL
5.3 Wall mounting (standard and option 1)
båÖäáëÜ
5.3
Wall mounting (standard and option 1)
In case of mounting on weight-bearing
wood structures
Use the enclosed wood screws and washers from the mounting kit for
mounting the unit on weight-bearing wood structures.
Setting up the unit
1. Remove the profile cover (A).
2. Position the two installation aids (B) at the foot (C) of the unit and
secure their position with adhesive tape.
CAUTION! The installation aids must be placed on top of each other
in such a way that their openings lie on top of each other.
NOTICE
Reduced image quality
Note the wall and ground conditions. If a carpet is present at the unit
installation site, it must be removed.
Each wall plug must support a tensile force of 700 N.
The wall construction must be suitable for the installation of the unit.
2x 8x80
2x Ø 8,4
NOTICE
Possible transport damage
Leave the center styrofoam packaging attached to the unit for
protection.