64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
183
Sirona Dental Systems GmbH
12 Startup, measurements and tests (for USA/Canada only)
Installation Manual ORTHOPHOS SL
12.6 Checking the radiation time
båÖäáëÜ
12.6
Checking the radiation time
Preparing the measurement
1. Attach the Mult-O-Meter sensor in the middle of the sensor unit (DCS-
side).
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
The unit moves to its starting position.
5. Call the Service menu and the Service routine S002.3 (see Service
Manual).
6. Use the arrow keys (A) in
selection field 1
to select the kV/mA level
60 kV/8 mA.
7. Use the arrow keys (A) in
selection field 2
to select the radiation time
0.5 s.
Performing measurements
➢ Initiate the radiation. Hold the release button pressed until the set
radiation time has expired.
Analyzing measurements
➢ Read the radiation time on the Mult-O-Meter.
The value for the radiation time displayed on the Mult-O-Meter
must correspond to the radiation time of 0.5s selected in the
service routine. The permissible tolerance is ± 10 %.
If the measured radiation time does
not
fall within the permissible
tolerance, replace the
tube assembly
(see Service Manual).
If the measured radiation time falls within the permissible
tolerance, finalize the measurement.
Concluding the measurement
1. Exit the service routine.
2. Switch the unit on via switch (A) (see also Operating Instructions).
A
Summary of Contents for ORTHOPHOS SL 2D
Page 2: ......