Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1 General information on unit adjustment and calibration
båÖäáëÜ
13.1.2.2.1
Navigation area
The navigation area contains a structure tree similar to the one you will
be familiar with from your Windows interface. The structure tree contains
all stages of the adjustment and calibration procedure you need to
complete in order to adjust and calibrate your system.
IMPORTANT: You must perform adjustment and calibration of the unit in
the order as described in the section “General information on unit
adjustment and calibration [ → 193]”.
Structure tree
Resetting adjustment/calibration and
data backup
●
"Reset Pan"
●
"Reset Ceph"
●
"Reset 3D"
2D adjustment (panoramic)
●
"Sensor adjustment"
●
"Diaphragm adjustment"
●
"Symmetrie adjustment"
●
"Sensor calibration (DCS)"
If a cephalometer is installed:
●
"Adjustment of Ceph primary diaphragm"
●
"Adjustment of Ceph secondary diaphragm"
●
"Adjustment of Ceph secondary diaphragm (Quickshot)"
●
"Adjustment of Ceph main beam direction"
●
"Adjustment of earplug alignment"
3D adjustment/calibration (volume)
●
"Sensor adjustment"
●
"Diaphragm adjustment"
●
"Sensor calibration"
●
"Diaphragm calibration"
●
"Geometry calibration"
●
"Dose measurement"