13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1 General information on unit adjustment and calibration
båÖäáëÜ
13.1.2.2.2
Coarse and precision adjustment
The pan adjustment menus
"Sensor adjustment"
and
"Diaphragm
adjustment"
and the Ceph adjustment menu
"Adjustment of Ceph
secondary diaphragm"
support coarse and precision adjustment
(precision adjustment is preset. Always try to use precision adjustment
first when adjusting the unit. In most cases, previous coarse adjustment
is not necessary.
A message window indicates that a coarse adjustment is required if
necessary.
To perform coarse adjustment, uncheck the
"Fine adjustment"
box and
complete the adjustment procedure in the same way as for precision
adjustment.
The correction procedure required for coarse adjustment is identical to
that for precision adjustment. The only difference between the two modes
is the size of the image field considered. Furthermore, there are fewer
auxiliary lines in the coarse adjustment mode.