13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
✔
The needle phantom is
not
in the bite block holder on the unit.
1. In the structure tree, under
"Pan"
, click on the
"Diaphragm
adjustment"
element (S030.2).
The
"Diaphragm adjustment"
menu is displayed in the action
area. The menu supports precision adjustment and coarse
adjustment (precision adjustment is pre-selected). Perform a
precision adjustment first. In most cases, previous coarse
adjustment is not necessary. [ → 203]
2. Establish receptivity [ → 209].
3. Create an exposure (60 kV / 3 mA; 0.3 s) [ → 209].
The adjustment values Sx and Sy are calculated automatically
from the exposure and entered in the text boxes of the
"Diaphragm adjustment"
menu.
The exposure is displayed in the exposure window.