64 95 142 D3632
236
D3632
.
031.01.02
.
02 03.2016
13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 237].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 209]
The status indicator that prefixes the
"Adjustment of Ceph
primary diaphragm"
element is checked.
6. Remove the ceph test phantom from the sensor unit cover.
7. Continue with the next stage of the adjustment procedure. [ → 238]
A
Adjustment OK
● The vertical pin must be
horizontally centered in the
exposed image section. A slight
vertical offset of the grid is
permissible.
● A uniform white border
surrounding the image on all sides
must be visible.
● Distance S3 must be approx. 60
mm.
B
Adjustment not OK
IMPORTANT: If S3is >70mm, please
contact the Sirona Customer Service
Center: