Sirona Dental Systems GmbH
11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
Checking the jumper position on board DX1
➢ Check the jumper positions of the sockets on board DX1:
X309 (DX91, ceph sensor) and
X503 (DX81, ceph sensor)
Without ceph
Jumper inside: Cephalometer not connected, i.e. connector not plugged
in.
Ceph left or Ceph right
Jumper outside: Cephalometer connected, i.e. connector plugged in.