11 Commissioning, demo mode and repacking
Installation Manual ORTHOPHOS SL
11.12 Service routines for commissioning and repacking
båÖäáëÜ
11.12.4.6
S018: Test step 7
NOTE: In order to execute this service routine, the occlusal bite block
function must be activated via service routine S017.18 [ → 147].
Setting the nominal angle for the occlusal bite block (as required by
the customer)
1. Call service routine S018.7.
The selected program group is displayed in selection field 1 (to
which the nominal value displayed in selection field 2 applies).
The nominal value (for the program group selected in selection
field 1) is displayed in selection field 2.
The currently measured angle of the bite block plate is displayed
and permanently updated in selection field 3
(possible deflection: -17° to +22°).
The display of patient symbol key 1 (G) indicates whether or not
the bite block is connected (see table).
Symbol on the control
panel
Status
Meaning
Patient symbol key 1 (G) visible
Occlusal bite block
connected
not visible
Occlusal bite block not
connected
Selection field
Parameter/Display
Range of values
1
Program groups
P1 - P12
BW1 - BW2
VOL1 - VOL3
2
Nominal angle of the bite block
plate
-13° to +18°
3
Measured deflection of the bite
block plate
-17° to +18°