13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
13.2.1.1.1
Automatic adjustment: Panoramic sensor (standard)
✔
The needle phantom is in the bite block holder on the unit [ → 210].
1. Open the calibration menu [ → 195].
2. At the bottom of the structure tree, click on the
"Sensor adjustment"
element (S010.1) under
"Pan"
.
The
"Sensor adjustment"
menu is displayed in the action area.
The menu supports precision adjustment and coarse adjustment
(precision adjustment is pre-selected). Perform a precision
adjustment first. In most cases, previous coarse adjustment is not
3. Establish receptivity [ → 209].
4. Create an exposure (60 kV / 3 mA ; 0.7 s). [ → 209]
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
the
"Sensor adjustment"
menu.
The exposure is displayed in the exposure window.