13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.1 General information on unit adjustment and calibration
Installation Manual ORTHOPHOS SL
1. Insert the geometry phantom (A) into the pan bite block holder (B) on
the unit and align the geometry phantom correctly with the screw (C).
2. Attach the clip (D) for better stability.
NOTICE
For correct calibration, the geometry phantom must be aligned
horizontally and vertically using a spirit level after insertion in the bite
block holder.