64 95 142 D3632
D3632
.
031.01.02
.
02 03.2016
73
Sirona Dental Systems GmbH
7 Installation: Remote control
Installation Manual ORTHOPHOS SL
7.3 Installation
båÖäáëÜ
7.3
Installation
The cable for the remote control can be installed on the surface or
beneath the surface (concealed). For
concealed installation
, the control
cable is drawn into the chassis from the rear. For
surface installation
, the
control cable is drawn into the chassis from underneath.
Preparations
1. Press carefully into the groove (A) from below with the slotted
screwdriver (do not lever).
2. Remove the lid from the chassis.
3. Hold the chassis against the wall in its mounting position and mark
the positions for the three drill holes with an awl.
4. Drill the holes and insert the wall anchors.
Concealed installation
1. Feed the control cable from the wall into the chassis through the rear
opening (B). The cable length between the wall outlet and the
stripped wire ends should be 250 mm (10").
2. Fasten the chassis firmly to the wall with the three screws (C).
A
Ø 6mm
Ø 1/4”
3x S 6
3x 4x30
C