Sirona Dental Systems GmbH
13 Unit adjustment and calibration
Installation Manual ORTHOPHOS SL
13.1 General information on unit adjustment and calibration
båÖäáëÜ
"Diaphragm adjustment"
(Pan - adjustment)
Diaphragm adjustment can usually be performed directly via precision
adjustment. Only in exceptional cases, e.g. if the exposed image area is
completely outside the image section in an image acquired with the
"Fine
adjustment"
setting (C), is it necessary to perform coarse adjustment
prior to precision adjustment (D).
On the exposure with coarse adjustment (D), the exposed area is just
barely visible at the right margin of the image field. Even in this extreme
case, an adjustment would still be possible.