5 Installation: X-ray unit
Sirona Dental Systems GmbH
5.4 Installing the floor stand (option 2)
Installation Manual ORTHOPHOS SL
5. Drill the fastening holes in the floor through the holes in the base
plate. Remove the drilling dust with a vacuum cleaner.
6. Slide the wall plugs (W) through the base plate and into the drillings.
7. Use the five screws and washers (X) to screw the base plate to the
floor.
X
W